| Catalog Number |
AB14W060040150 |
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The physician used a nanocross elite with a non-medtronic inflation device for the treatment of a severely calcified lesion in the patients distal superficial femoral artery (sfa).The lesion exhibited 90% stenosis.Slight tortuosity was reported.The ifu was followed and the device was prepped without issue.It is reported that during balloon inflation a longitudinal balloon burst occurred at a pressure of 8 atms.An additional 6x40 balloon was opened and used to successfully complete the procedure.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|