• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. FIBERTAPE TENDON COMPRESSION BRIDGE KIT; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FIBERTAPE TENDON COMPRESSION BRIDGE KIT; SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE Back to Search Results
Model Number FIBERTAPE® TENDON COMPRESSION BRIDGE KIT
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/03/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device is expected but was not returned for evaluation.The cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during a total shoulder arthroplasty procedure as the surgeon was attempting to pull the tails of the tape on the ar-7297 fibertape compression bridge kit (that were loaded into the nitinol loop) through the looped ends of the tape, the nitinol loop broke before the ends of tape were passed.The broken black plastic piece and nitnol loop were fully retrieved and tossed from the sterile field.The procedure was completed as planned.The rep stated that the surgeon performs x-rays after every procedure.The x-ray taken revealed a small metal piece remained in the patient.The patient was still under anesthesia and in the operating room.The surgeon reopened the incision and extended by approximately an inch in order to remove the broken fragment.After inspecting the broken metal piece, the rep confirmed that the piece that fell into the incision was the small metal rod within the black plastic piece and nitnol loop.Additional information received 07/10/2018: the rep stated that the procedure was completed by loading the ends of the tape into a free needle, and proceeded to pass the ends through the tape loops.Additional information received 7/26/2018 via facility medwatch (4900240000-2018-8091): foreign body was retrieved with flouro guidance.Routine flat plate films were done at the end of the second procedure.No retained object confirmed.Surgeon notified family before starting foreign body removal procedure and after it was completed.
 
Additional Manufacturer Narrative
The evaluation revealed the nitinol wire had pulled out of the ferrule.The ferrule met all dimensional specifications as received and displayed deep crimp markings as expected.The nitinol wire was not received with the device, so it could not be determined if the wire placement within the ferrule was correct.The retention dowel displayed tool marks along the shaft consistent with attempts to retrieve the device during surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERTAPE TENDON COMPRESSION BRIDGE KIT
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYETHYLENE
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key7734409
Report Number1220246-2018-00574
Device Sequence Number806077
Product Code GAT
UDI-Device Identifier00888867289086
UDI-Public00888867289086
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K122374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model NumberFIBERTAPE® TENDON COMPRESSION BRIDGE KIT
Device Catalogue NumberAR-7297
Device Lot Number10197428
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2018
Initial Date Received by Manufacturer 07/10/2018
Supplement Date Received by Manufacturer07/10/2018
Initial Report FDA Received Date07/31/2018
Supplement Report FDA Received Date08/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-