Model Number 174021
Medical Device Problem Codes Break (1069); Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 04/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of one device.Visual inspection of the instrument noted no visual abnormalities.During firing, the tacks seated properly, but handle did not retract after firing.The leaf spring was loose on handle.The unit was opened as instructed by engineering and it was noted the leaf spring was detached and rolled.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.However, an assembly error was identified during product analysis.Root cause of the reported condition is due to an excessive amount of adhesive being applied to the spring retainer during the assembly process.A process improvement has been initiated to prevent this condition from recurring.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device was in small pieces.There was no patient involvement.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIFIRE VERSATACK
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
MDR Report Key7734753
Report Number2647580-2018-03826
Device Sequence Number1398907
Product Code GCJ
UDI-Device Identifier20884521070568
UDI-Public20884521070568
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K912097
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number174021
Device Catalogue Number174021
Device Lot NumberP7C0904X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/19/2018
Initial Date Received by Manufacturer 07/06/2018
Supplement Date Received by Manufacturer08/08/2018
08/08/2018
Initial Report FDA Received Date07/31/2018
Supplement Report FDA Received Date10/03/2018
10/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-