| Model Number |
174021 |
| Medical Device Problem Codes |
Break (1069); Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.Visual inspection of the instrument noted no visual abnormalities.During firing, the tacks seated properly, but handle did not retract after firing.The leaf spring was loose on handle.The unit was opened as instructed by engineering and it was noted the leaf spring was detached and rolled.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.However, an assembly error was identified during product analysis.Root cause of the reported condition is due to an excessive amount of adhesive being applied to the spring retainer during the assembly process.A process improvement has been initiated to prevent this condition from recurring.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, the device was in small pieces.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|