| Model Number |
4401204S |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis.Device history record (dhr) review was conducted and the product met quality requirements for product acceptance.Additional information is pending and will be submitted within 30 days upon receipt.
|
| |
|
Event or Problem Description
|
|
During use, it was reported that the power flex pro ruptured at six atmosphere (6 atm) which was within its rated balloon pressure.There was no patient injury.The lesion was the iliac artery with calcification.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received and section was updated below.During use, it was reported that the power flex pro ruptured at six atmosphere (6 atm) which was within its rated balloon pressure.There was no patient injury.The lesion was the iliac artery with calcification.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The device is not available for analysis.No other information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
During use, it was reported that the 12x40cm 80cm powerflex pro balloon catheter (bc) ruptured at six atmosphere (6 atm) which was within its rated balloon pressure.There was no patient injury.The lesion was the iliac artery with calcification.There were no difficulties removing the product from the hoop, the protective balloon cover, the stylet or any of the sterile packaging components.There were no kinks or other damages noted prior to inserting the product the product into the patient.The device is not available for analysis.No other information was provided.The device was not returned for analysis.A device history record (dhr) review of lot 17703241 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause of the burst could not be determined.Based on the limited information available for review, vessel characteristics (calcification) and procedural/handling factors may have contributed to the reported event since calcified/resistant lesions can cause damage to a balloon.According to the instructions for use, which is not intended as a mitigation, ¿the rated burst pressure is based on the result of in vitro testing.At least 99.9% of the balloons (with a 95% confidence level) will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr review nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective/preventive action will be taken at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
During use, it was reported that the power flex pro ruptured at six atmosphere (6 atm) which was within its rated balloon pressure.There was no patient injury.The lesion was the iliac artery with calcification.There is no information about handling, prep or condition of the device when it was removed from the packaging.The device was not returned for analysis.A device history record (dhr) review of lot 17703241 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/below rbp (peripheral)¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.Based on the limited information available for review, vessel characteristics of calcification may have contributed to the reported event.Per the instructions for use (ifu), which is not intended as a mitigation, per the instructions for use ¿balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in-vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the dhr nor the information available suggests a design or manufacturing related cause for the reported event; therefore, no corrective/preventive action will be taken.
|
| |
|