MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8637-40
Device Problems Suction Problem (2170); Infusion or Flow Problem (2964)
Patient Problems Therapeutic Effects, Unexpected (2099); No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2018
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id: 8780, serial# (b)(4), implanted: (b)(6) 2018, explanted: (b)(6) 2018, product type catheter section d information references the main component of the system.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), (b)(6), (b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient via a manufacturer's representative (rep).The patient was receiving 500mcg/ml baclofen at 1 00mcg/day via an implantable infusion pump for an unknown indication for use.It was reported that the patient was not getting sufficient therapy like they did during the trial.It was reported that a dye study was performed and the healthcare professional (hcp) was unable to aspirate the catheter.The date of the dye study was unknown.It was reported that the catheter was replaced on (b)(6) 2018 and it was extremely difficult to get cerebrospinal fluid back and the hcp had to use x-ray to verify.It was reported that the issue was resolved at the time of the report.It was reported that the patient's status at the time of the report was noted as "alive-no injury." no further complications were anticipated/reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key7735338
MDR Text Key115564715
Report Number3004209178-2018-17028
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00643169530126
UDI-Public00643169530126
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2019
Device Model Number8637-40
Device Catalogue Number8637-40
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/27/2018
Initial Date FDA Received07/31/2018
Supplement Dates Manufacturer Received08/08/2018
Supplement Dates FDA Received10/04/2018
Date Device Manufactured05/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age62 YR
-
-