INTUITIVE SURGICAL, INC ENDOWRIST; MEGA SUTURECUT ND Back to Search Results
Model Number 470309-12
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/02/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure of ¿broken cable¿.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.Device history record (dhr) review-dhr review for the device(s) involved with the reported event has been completed.No non-conformances were identified to be related to this complaint.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
 
Event or Problem Description
It was reported that during a da-vinci assisted surgical procedure, the mega suturecut needle driver had a broken cable.There was no report of fragments falling inside the patient.The planned surgical procedure was completed and no patient harm, adverse outcome, or injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Common Device Name
MEGA SUTURECUT ND
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
950 kifer rd.
sunnyvale CA 94086
MDR Report Key7735778
Report Number2955842-2018-10418
Device Sequence Number2643543
Product Code NAY
UDI-Device Identifier00886874112434
UDI-Public(01)00886874112434(10)N10170821
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K131861
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470309-12
Device Lot NumberN10170821 0228
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2018
Initial Report FDA Received Date07/31/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-