| Model Number |
470309-12 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/02/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the instrument involved with this complaint and completed the device evaluation.Failure analysis confirmed the customer reported failure of ¿broken cable¿.The instrument was found to have a broken pitch cable at the distal clevis hub.The broken cable segment that contains the crimp was still installed in the clevis.Device history record (dhr) review-dhr review for the device(s) involved with the reported event has been completed.No non-conformances were identified to be related to this complaint.The customer reported complaint does not itself constitute an mdr reportable event; however, the broken pitch cable found during failure analysis could cause or contribute to an adverse event if the malfunction were to recur.
|
| |
|
Event or Problem Description
|
|
It was reported that during a da-vinci assisted surgical procedure, the mega suturecut needle driver had a broken cable.There was no report of fragments falling inside the patient.The planned surgical procedure was completed and no patient harm, adverse outcome, or injury was reported.
|
| |
|