(b)(4).The device history records were reviewed and the manufacturing criteria was met prior to the release of this lot.The product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
|
(b)(4).Date sent to the fda: 09/18/2018 evaluation: a sealed box, an unopened box with an empty opened foil, and unopened samples of product were returned for analysis.During the visual inspection of the samples, the information of product code, lot number and graphic on the label overwrap and paper lid printed was compared with system database.In addition, a characterization was performed.The characterization of the sample was conducted, and the following was observed: the sample was measured in needle diameter, point type, needle type, needle radius/curvature and angle belong to product code d5264.Also, the suture was measured in diameter suture and length and the structure of the strand belong to vicryl suture diameter 6-0¿and it is correct to the product code d5264.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.However, when the code was reviewed on the ethicon web product catalog, it was found to have a different product on the photo and the characteristics don¿t correspond to the physical samples received.The product show at the web page belongs to a different product.Per the condition of the sample received and the information of the customer, the difference on the homepage corresponds to the reported complaint.However, the samples received meet the finish good requirements.
|