| Model Number |
LF4200 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
08/17/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation summary: one device was received for evaluation.This device had been used in the treatment or diagnosis of a patient.Reported and outside expiration date at time of reported incident.The returned product did not meet specification as received.The visual inspection found the knife was trapped and contained within the jaws.The jaws would not open or close due to the trapped knife.The reported condition was confirmed.The knife is trapped and contained within the jaws.The knife could not advance forward or retract.Knife trap happens when the blade is extended and the jaws are not completely closed.This allows the knife track to open too wide and the blade moves out of its track.As a safety measure, the knife must be retracted in order to open the jaws.The tissue in the webbing may have prevented the knife from retracting far enough to allow the jaws to open.More frequent cleaning could have also reduced the difficulty activating the knife.The investigation identified the root cause of the reported event to be knife trap due to user error.The instructions for use (ifu) sates, confirm that the jaws have reached the closed position and are locked (the handle is latched) before activating the cutter.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, the device had a cutting/knife issue, the knife blade did not advance.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|