JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS SYMFONY TORIC; MULTIFOCAL IOLS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown, not provided.If explanted, give date: not applicable, as the lens remains implanted.It was indicated that the device remains implanted and not returning for evaluation; therefore a failure analysis of the complaint device cannot be completed.A review of the device/lot history record, labeling, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a zxt150 17.5 diopter lens was implanted in the patient's right eye (od) on (b)(6) 2018.Post-operatively, the lens rotated 12 degrees.The patient still has more residual cylinder than anticipated along with a myopic outcome.The patient is wearing glasses and is not happy about it.Reportedly, the patient doesn't wish to do another surgery at this time.The patient's pre-operative refraction was -2.50 x 1.50 x 168 degrees and post-operative refraction was -1.00 x 1.00 x 015 degrees, including an uncorrected visual acuity (ucva) of 20/30 and a best corrected visual acuity (bcva) of 20/20.No additional information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7736246
Report Number9614546-2018-00735
Device Sequence Number11017673
Product Code POE
UDI-Device Identifier05050474605152
UDI-Public(01)05050474605152(17)220702
Combination Product (Y/N)Y
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/02/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U175
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/02/2018
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date07/31/2018
Supplement Report FDA Received Date11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-