Model Number X SERIES |
Device Problem
Failure to Read Input Signal (1581)
|
Patient Problem
Ventricular Tachycardia (2132)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device would not go into sync mode.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation evaluated the device and the device performed to specification.Review of the device activity logs indicate sync markers were detected in pads mode when the criteria was met.Further review of the activity logs also indicated that the device was never switched from pads view to a different ecg lead during the event.The device was put through extensive testing which included functional testing and it met zoll's specifications.Based on the evidence presented in the clinical file, the investigation is being closed as the device met specification.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
|