ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Device Problem Failure to Read Input Signal (1581)
Patient Problem Ventricular Tachycardia (2132)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device would not go into sync mode.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
Zoll medical corporation evaluated the device and the device performed to specification.Review of the device activity logs indicate sync markers were detected in pads mode when the criteria was met.Further review of the activity logs also indicated that the device was never switched from pads view to a different ecg lead during the event.The device was put through extensive testing which included functional testing and it met zoll's specifications.Based on the evidence presented in the clinical file, the investigation is being closed as the device met specification.The device was recertified and returned to the customer.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key7736354
MDR Text Key115647593
Report Number1220908-2018-02138
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946004354
UDI-Public00847946004354
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112432/P160
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/10/2018
Initial Date FDA Received07/31/2018
Supplement Dates Manufacturer Received07/10/2018
Supplement Dates FDA Received08/24/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-