A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
|
It was reported that during treatment with a prismaflex m150 set, an air alarm was generated.A loose connection was identified at the access line/dialyzer inlet connection.The connection was tightened properly; however, treatment was terminated as the amount of air in the extracorporeal circuit (ec) was determined as ¿so much¿.The blood in the extracorporeal circuit was partially returned.A blood loss of approx.100 ml occurred.It was reported a ¿new system was installed¿ and treatment was continued.There was no patient injury or medical intervention associated with this event.No additional information is available.
|