COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Device Problem Overheating of Device (1437)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer states that the enteral feeding pump heats up.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
An evaluation of the kangaroo epump was performed and the customer states ¿the enteral feeding pump heats up.¿ the unit was triaged and the customer¿s reported condition was confirmed.Internal inspection and investigation reveals thermal damage on components c95 and c104 in the ac adapter circuit.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
15 hampshire street
mansfield MA 02048
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7736705
MDR Text Key115677436
Report Number1282497-2018-00639
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006614
UDI-Public10884521006614
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-