(b)(4).Evaluation statement: this complaint device was received and evaluated.Visual and functional investigation confirms that jaw is not operational.It appears as though the spring that facilitates the actuating of the jaw is broken contributing to this failure.This complaint can be confirmed.This failure is typical of excessive abuse of the device over time which would eventually lead to jaws failing.A batch record review has been conducted for this lot of pieces that was manufactured in 2014 and the results indicate that in this batch of product there were no anomalies or discrepancies in the manufacture of this lot and therefore there is no internally assignable cause for the reported problem.Further, a review into the depuy mitek complaints system shows no other complaints for this lot of devices.At this point in time, no further action is warranted.However, depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|