MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET 360 NANO 3 MM X 6 CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET 360 NANO 3 MM X 6 CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0035443060
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device is not available.
 
Event or Problem Description
It was reported that during embolization procedure with the subject coil, ¿the doctor pressed the sheath against the hub of the microcatheter too much so the tip of the coil broke¿.There were no clinical consequences to the patient.
 
Additional Manufacturer Narrative
The device history record review confirms that the device met all material, assembly and performance specifications.The subject device was not returned for analysis; therefore, physical as well as a functional testing could not be performed.In the case of this complaint it is possible that the main coil broke due to excessive force applied on coil at hub during insertion.Based on the information currently available the exact cause for the reported main coil break cannot be determined.Expiration date: updated.Manufacturing date: updated.
 
Event or Problem Description
It was reported that during embolization procedure with the subject coil, ¿the doctor pressed the sheath against the hub of the microcatheter too much so the tip of the coil broke¿.There were no clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 3 MM X 6 CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7737478
Report Number3008881809-2018-00327
Device Sequence Number2643933
Product Code HCG
UDI-Device Identifier07613252600329
UDI-Public07613252600329
Combination Product (Y/N)N
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2021
Device Catalogue NumberM0035443060
Device Lot Number20453184
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/05/2018
Supplement Date Received by Manufacturer08/13/2018
Initial Report FDA Received Date07/31/2018
Supplement Report FDA Received Date08/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-