• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH CORTICAL FIX X-TAB 7X40MM TI; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH CORTICAL FIX X-TAB 7X40MM TI; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 186770140
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Further to (b)(4), on (b)(6) 2018, when the patient visited the hospital due to a lower back pain, the breakage of the rod (196789600) was found under the x-ray between l4 and l5 (right), and between l5 and s (left).On (b)(6) 2018, the re-operation was rescheduled due to the patient¿s high crp (c-reactive protein) level.On (b)(6) 2018, the re-operation was performed to find the cause of the patient¿s high crp (c-reactive protein) level, considering the suspected infections.The procedures were conducted in the following way.Screws at l4, l5 and s1 (right) were extracted and new screws were inserted at l5 and s1 to connect with upper rods using a connector.A screw at l5 (left) was extracted to connect with upper rods using a connector.Si set screws at l4 and over (both sides) were replaced with new ones, and rods at far end were also replaced with new ones.Also, the following were commented by the surgeon.Loosening of all the screws were observed when extracted torque was not properly applied to si set screws when extracted.Black-colored substances were seen around the screw and the rod, which might have been caused by the metal corrosion.The patient¿s crp level might have increased due to armd caused by the metal corrosion.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).No product failure was alleged against this device and no failures were observed during evaluation.No root cause analysis is necessary at this time as no product failures have been identified during the course of this investigation.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORTICAL FIX X-TAB 7X40MM TI
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key7737485
Report Number1526439-2018-50724
Device Sequence Number805971
Product Code NKB
UDI-Device Identifier10705034465784
UDI-Public(01)10705034465784
Combination Product (Y/N)N
PMA/510(K) Number
K160904
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2018
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number186770140
Device Catalogue Number186770140
Device Lot NumberTBJXI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Initial Date Received by Manufacturer 07/02/2018
Supplement Date Received by Manufacturer08/23/2018
Initial Report FDA Received Date07/31/2018
Supplement Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-