Model Number ZXT150
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Codes Halo (2227); No Code Available (3191)
Date of Event 11/09/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If explanted, give date: not applicable as to date the lens remains implanted, however the surgeon is planning to explant.(b)(4).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the zxt150 +11.5 diopter intraocular lens (iol) was implanted in the patient's left eye (os) on (b)(6) 2017.Reportedly, the lens was repositioned on (b)(6) 2017.The patient visited the doctor on (b)(6) 2017, and complained of pseudophakia ou (both eyes).The patient is frustrated that he cannot read, which is affecting his work.He bought +1.25 otc (over-the-counter) glasses with no help.Intermediate vision has improved.Distance vision os stable; fuzzy od (right eye).Also noted halos.Uncorrected visual acuity os: 20/25+2 distance; j2 near.Additional information was received and it was learned that the surgeon is planning to explant the lens from the left eye; no date set.No further information was provided.This report is to record the lens in the left eye.A separate mdr will be filed for the lens in the right eye.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site as to date it remains implanted.Therefore product testing could not be performed and the customer''s reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7737493
Report Number9614546-2018-00738
Device Sequence Number19062718
Product Code POE
UDI-Device Identifier05050474605039
UDI-Public(01)05050474605039(17)220620
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/20/2022
Device Model NumberZXT150
Device Catalogue NumberZXT150U115
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/06/2018
Supplement Date Received by Manufacturer09/04/2018
10/25/2020
Initial Report FDA Received Date07/31/2018
Supplement Report FDA Received Date09/14/2018
11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-