Model Number RT265
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 04/20/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint rt265 infant dual heated evaqua2 breathing circuit was not returned to fisher & paykel healthcare in (b)(4) for evaluation.Several attempts to obtain further information and the complaint device from the healthcare facility were made.Without the return of the complaint device we are not able to determine what may have caused the reported event described by the customer.All rt265 infant dual heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.A circuit with this damage would not have passed our testing.Our user instructions that accompany the rt265 state the following: check all connections are tight before use.Perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient.Set appropriate ventilator alarms.
 
Event Description
A healthcare facility in (b)(6) reported via a distributor that the rt265 infant dual heated evaqua2 breathing circuit is circuit is "not working properly/broken".The customer further reported that the circuit failed the initial ventilator leak test prior to patient use.This was found prior to patient use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, auckland 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, auckland 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key7737498
MDR Text Key115670739
Report Number9611451-2018-00636
Device Sequence Number1
Product Code BZE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT265
Device Catalogue NumberRT265
Device Lot NumberNOT PROVIDED
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/03/2018
Initial Date FDA Received07/31/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-