• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (PRINGY) JUVEDERM (VOLUME/CONCENTRATION/ADDITIVE UNKNOWN); IMPLANT, DERMAL, FOR AESTHETIC USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (PRINGY) JUVEDERM (VOLUME/CONCENTRATION/ADDITIVE UNKNOWN); IMPLANT, DERMAL, FOR AESTHETIC USE Back to Search Results
Catalog Number UNK JUVEDERM-CONC/ADD
Device Problems Gel Leak (1267); Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Further information from the reporter regarding event, product, and patient details has been requested.No additional information is available at this time.Device labeling addresses the reported event(s) as follows: method of use-posology ¿ juvéderm® voluma¿ with lidocaine is to be used as supplied.Modification or use of the product outside the directions for use may adversely impact the sterility, homogeneity and performance of the product and it can therefore no longer be assured.¿ remove tip cap by pulling it straight off the syringe.Then firmly push the needle provided in the box into the syringe, screwing it gently clockwise.Twist once more until it is fully locked.Next, remove the protective cap by holding the body of the syringe in one hand, the protective cap in the other, and pulling the two hands in opposite directions.Prior to injecting, depress the plunger rod until the product flows out of the needle.Inject slowly and apply the least amount of pressure necessary.If the needle is blocked, do not increase the pressure on the plunger rod.Instead, stop the injection and replace the needle.Failure to comply with these precautions could cause a disengagement of the needle and/or product leakage at luer-lock level and/or increase the risk of vascular compromise.
 
Event Description
Healthcare professional reported the physician went to plunge the needle [unspecified juvéderm®] exploded out into the hands, physician was completely shocked, the needle was on securely, there was 0.3mls left and the patient wanted the physician to try again when the physician reinserted the needle and then it would plunge out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JUVEDERM (VOLUME/CONCENTRATION/ADDITIVE UNKNOWN)
Type of Device
IMPLANT, DERMAL, FOR AESTHETIC USE
Manufacturer (Section D)
ALLERGAN (PRINGY)
route de promery
zone artisanale de pre-mairy
pringy 74370
FR  74370
Manufacturer (Section G)
ALLERGAN (PRINGY)
route de promery
zone artisanale de pre-mairy
pringy 74370
FR   74370
Manufacturer Contact
suzanne wojcik
301 w howard lane
suite 100
austin, TX 78753
7372473605
MDR Report Key7737751
MDR Text Key115649264
Report Number3005113652-2018-00977
Device Sequence Number1
Product Code LMH
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P050047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 07/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK JUVEDERM-CONC/ADD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-