Model Number V60
Device Problem Disconnection (1171)
Patient Problems Death (1802); Pneumonia (2011)
Event Date 11/10/2017
Event Type  Death  
Manufacturer Narrative
(b)(6).Udi: (b)(4).The customer reported that a patient on a v60 ventilator passed away after the mask became disconnected from the patient.The ventilator audibly alarmed, however, the nurse did not notice that the mask had come off the patient until approximately ten minutes later.The ventilator was evaluated by philips and no abnormality was found.The ventilator error logs were reviewed and no hardware issues were found.The unit was returned to service and will not be returned.
 
Event Description
The customer reported that a patient with pneumonia on a v60 ventilator passed away after the mask became disconnected from the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
RICA
2271 cosmos court
carlsbad CA 92011
Manufacturer Contact
lisa cardenas
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key7739031
MDR Text Key115684072
Report Number2031642-2018-01526
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838020054
UDI-Public00884838020054
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/06/2018
Initial Date FDA Received08/01/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN MASK AND CIRCUIT.
Patient Outcome(s) Death;
Patient Age74 YR
-
-