MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD SURGITIE; SUTURE, ABSORBABLE, NATURAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD SURGITIE; SUTURE, ABSORBABLE, NATURAL Back to Search Results
Model Number EL-20-L
Medical Device Problem Code Difficult to Open or Close (2921)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 07/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter during a laparoscopic appendectomy, there was a strong resistance when pulling the string, so the loop could not be narrowed and the ligation could not be completed.The procedure was completed with another device.There was no patient harm.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted the tubing was fractured.The loop was broken off at the knot.The loop appeared to have been fully cinched.As no sealed samples were returned, functional testing could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURGITIE
Common Device Name
SUTURE, ABSORBABLE, NATURAL
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7739108
Report Number9612501-2018-01390
Device Sequence Number16404063
Product Code GAL
UDI-Device Identifier10884521099647
UDI-Public10884521099647
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K904588
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2023
Device Model NumberEL-20-L
Device Catalogue NumberEL-20-L
Device Lot NumberJ8C1955X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2018
Initial Date Received by Manufacturer 07/17/2018
Supplement Date Received by Manufacturer08/08/2018
03/18/2019
Initial Report FDA Received Date08/01/2018
Supplement Report FDA Received Date10/02/2018
03/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-