| Model Number |
EL-20-L |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter during a laparoscopic appendectomy, there was a strong resistance when pulling the string, so the loop could not be narrowed and the ligation could not be completed.The procedure was completed with another device.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.A visual inspection of the returned product noted the tubing was fractured.The loop was broken off at the knot.The loop appeared to have been fully cinched.As no sealed samples were returned, functional testing could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|