MAUDE Adverse Event Report: AV-TEMECULA-CT SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING STENT SYSTEM Back to Search Results
Catalog Number S-55-120-120-P6
Medical Device Problem Codes Stretched (1601); Difficult or Delayed Activation (2577)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents.The investigation was unable to determine a cause for the reported deployment difficulty.The elongation was due to operational context.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified, superficial femoral artery.Pre-dilatation was performed.During deployment of the 5.5 x 120 mm supera stent system, the stent would not release, the stent then elongated, doubled in length and the stent released.The stent was deployed in the target lesion and unhealthy tissue.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7739379
Report Number2024168-2018-05935
Device Sequence Number13857646
Product Code NIP
UDI-Device Identifier08717648226144
UDI-Public08717648226144
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2019
Device Catalogue NumberS-55-120-120-P6
Device Lot Number8012961
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2018
Initial Report FDA Received Date08/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-