| Catalog Number |
364979 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that when transferring urine from urine cups with integrated transfer device into at bd vacutainer® urinalysis urine tubes, bubbles formed in the conical of the tubes which caused the instrument to read them erroneously as false positive red blood cells.The doctor recalled the patients to get retested.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: sex: female.Event attributed to: other.Device single use?: no.Device returned to manufacture: yes.Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for erroneous results with the incident lot was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa and potential causes have been identified.As a result, corrective actions have been established and are in the process of being implemented.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for erroneous results with the incident lot was not observed.Further investigation activities have been conducted through a capa and the most likely root cause has been identified.As a result, corrective actions and procedures are being implemented to mitigate further occurrences.Root cause description: a capa was conducted to document further investigation and root cause analysis relating to this issue.The investigation has identified the most likely root causes and corrective actions are in the process of being implemented.Based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue.The investigation has identified potential root cause(s) for this issue and corrective actions are in the process of being implemented.
|
| |
|
Event or Problem Description
|
|
It was reported that when transferring urine from urine cups with integrated transfer device into at bd vacutainer® urinalysis urine tubes, bubbles formed in the conical of the tubes which caused the instrument to read them erroneously as false positive red blood cells.The doctor recalled the patients to get retested.
|
| |
|