Model Number 72201919S
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/25/2018
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the image was lost during the case.There was a backup device available to complete the procedure.No patient injuries reported.
 
Additional Manufacturer Narrative
The device was received for evaluation.A visual inspection was performed on the product and no issue was observed.There was a relationship found between the returned device and the reported incident.Ccu failed functional testing with intermittent live dvi video output after 2 hour burn-in.Cause of dvi video output malfunction is a defective electronic component on the video processor pcb.Product passed functional testing with a known good video processor pcb installed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVICE CCU HIGH DEF 560P 2NDGEN
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
MDR Report Key7739941
Report Number1643264-2018-00559
Device Sequence Number13857637
Product Code GCJ
UDI-Device Identifier03596010645517
UDI-Public(01)03596010645517
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K070266
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72201919S
Device Catalogue Number72201919S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2018
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer08/16/2018
Initial Report FDA Received Date08/01/2018
Supplement Report FDA Received Date08/18/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-