• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COAGUCHEK XS PRO KIT; PROTHROMBIN TIME MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COAGUCHEK XS PRO KIT; PROTHROMBIN TIME MONITOR Back to Search Results
Catalog Number 05530199160
Medical Device Problem Code Unauthorized Access to Computer System (3025)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/01/2018
Type of Reportable Event Malfunction
Event or Problem Description
An internal investigation has determined that coaguchek xs pro meters are vulnerable to a cybersecurity or data breach.This was discovered during roche internal investigations in conjunction with a cybersecurity consultant.The internal investigation determined that significant product and cybersecurity knowledge is needed to exploit these vulnerabilities.The consultant was able to reverse engineer the service password generator.The consultant was also able to execute code through a hidden command.The investigation determined the following: -deletion of protected health information may be possible, -modification of protected health information may be possible, -changing the units of measure may be possible, -the vulnerabilities can be exploited independently or in combination.In the worst case scenario, the medical risk is high.The probability of this type of event occurring is low (criminal intent, theft of the device, entering the firewall, etc.).The likelihood of occurrence of a successful attack is deemed low to theoretical with installations secured in accordance with roche cyber security recommendations.There have been no reports of actual occurrence in the field.
 
Additional Manufacturer Narrative
Successful exploitation of the vulnerabilities may have the following consequences: protected health information and other personally identifiable information on the devices may be disclosed and therefore cause violation of data protection laws.Crashed or corrupted devices may disrupt operation in the healthcare delivery organization.Manipulated result data may lead to wrong diagnosis and treatment of patients.The attack complexity has been assessed as high, as attackers with criminal intent would need to physically or logically compromise the healthcare network, perform extensive reverse engineering to disclose the vulnerabilities and develop targeted exploit code based on in-depth technical knowledge about the roche products to successfully exploit the vulnerabilities.Likelihood of occurrence of a successful attack is deemed low to theoretical with installations secured in accordance with roche cyber security recommendations.The attack can only be detected by chance.Since it needs criminal intent an attacker will try to hide his activities as much as possible.The product itself is not defective.The patient results would only be affected (by changing value/unit etc.) due to intentionally induced malfunctions by third parties.To be able to affect patient results the device needs to be manipulated/misused purposefully in an illegal way.If an attacker succeeds in altering patient results either by a direct attack or by implementing malicious software, a subsequent medical risk for point of care diagnostics cannot be excluded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PRO KIT
Common Device Name
PROTHROMBIN TIME MONITOR
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7740438
Report Number1823260-2018-02557
Device Sequence Number16493704
Product Code GJS
UDI-Device Identifier00365702398108
UDI-Public365702398108
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K062925
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number05530199160
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/02/2018
Supplement Date Received by Manufacturer07/02/2018
07/02/2018
Initial Report FDA Received Date08/01/2018
Supplement Report FDA Received Date08/20/2018
09/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-