ZYNO MEDICAL, LLC ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM; LARGE VOLUME INFUSION PUMP Back to Search Results
Model Number Z-800
Medical Device Problem Codes Excess Flow or Over-Infusion (1311); Improper Flow or Infusion (2954); Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zyno medical is waiting for the arrival of the device.
 
Event or Problem Description
A user facility representative reported an over-infusion issue to zyno medical on (b)(6) 2018: "the pump was set for 30 min and infused the patient in 13 min.The patient was not harmed and is okay".No information on solution being infused at the time of the event.The issue was first noticed on (b)(6) 2018 and the affected pump had not been used since then.
 
Event or Problem Description
This is a follow-up for the initially filed mdr (3006575795-2018-00057).
 
Additional Manufacturer Narrative
The pump was evaluated on 08/21/2018.The reported over-infusion issue was not confirmed.The pump was found to be under-infusing with a flow rate accuracy of -5.96%, which was outside of +/-5% specification.The pump was tested, recalibrated and met full specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZYNO MEDICAL Z-800 INFUSION PUMP SYSTEM
Common Device Name
LARGE VOLUME INFUSION PUMP
Manufacturer (Section D)
ZYNO MEDICAL, LLC
177 pine street
natick MA 01760
MDR Report Key7740958
Report Number3006575795-2018-00057
Device Sequence Number16423591
Product Code FRN
UDI-Device Identifier00814371020006
UDI-Public00814371020006
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K130690
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZ-800
Device Catalogue NumberZ-800
Device Lot Number20130227-SH
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/26/2018
Supplement Date Received by Manufacturer08/21/2018
Initial Report FDA Received Date08/01/2018
Supplement Report FDA Received Date09/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-