COOK INC TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER TRAY; DQY CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number C-UTLMY-501J-RSC-CCT
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 07/09/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pma/510(k) #: preamendment.(b)(4).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported a triple lumen polyurethane central venous catheter (cvc) was placed in the right internal jugular on (b)(6) 2018 for a pericardial effusion.The blue hub was found cracked on (b)(6) 2018.The patient required an additional procedure to remove and replace the device."the customer mentioned the issue of over tightening and it was pointed out by the hospital staff that is always the blue lumen that is cracking." the device was used to deliver continuous, scheduled and one time dose medications.The line was flushed with heparin as needed.As reported, the patient did not experience any other adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information has previously been reported, is unknown or is unavailable.Device evaluation the reported device was not returned to cook inc.For evaluation nor were unused devices from the same lot.Investigation a document-based investigation reviewed the following: instructions for use, quality control, specifications, complaint history, device history records and manufacturing instructions.A review of the device history record of the finished product shows no nonconforming events that could contribute to this failure mode.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.A complaint history search revealed that there were no other reported complaints for this lot number.Based on the information provided and the results of our investigation, a definitive root cause cannot be determined at this time.Per the [quality engineering] risk assessment, no further risk reduction action is required however appropriate measures have been initiated to investigate this failure mode.Monitoring will continue to be performed for similar complaints.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
See h10.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.- attachment: [mw5080167.Pdf].
 
Event or Problem Description
A sus voluntary report, mw5080167, was received on 25oct2018 and reconciled with this case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIPLE LUMEN POLYURETHANE CENTRAL VENOUS CATHETER TRAY
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7741037
Report Number1820334-2018-02226
Device Sequence Number1050207
Product Code DQY
UDI-Device Identifier00827002269453
UDI-Public(01)00827002269453(17)191101(10)8564700
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/28/2018
Report Date (Section F) 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2019
Device Catalogue NumberC-UTLMY-501J-RSC-CCT
Device Lot Number8564700
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Initial Date Received by Manufacturer 07/11/2018
Supplement Date Received by Manufacturer10/25/2018
11/28/2018
Initial Report FDA Received Date08/01/2018
Supplement Report FDA Received Date11/21/2018
12/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
A CAREFUSION SMARTSITE 2000E
Outcome Attributed to Adverse Event Required Intervention;
Patient Age8 YR
Date Report Sent to FDA09/25/2018
-
-