Catalog Number 1125350-15
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4) medical devices: guide wire: pt graphix.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation determined that the reported failure to advance, stent dislodgement, difficult to remove and foreign body in patient appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the heavily tortuous saphenous vein graft.The 3.5 x 15 mm xience alpine stent delivery system (sds) was advanced into the patient anatomy and an attempt was made to cross through a previously implanted unspecified stent to get to the target lesion that was distal to the implanted stent.The sds was unable to cross and the guide catheter was advanced further into the vessel.An attempt was made to remove the sds; however, resistance was noted with the anatomy and all devices were removed as a single unit.Upon removal, it was noted that the stent was no longer on the sds.The stent could not be located on the devices or in the patient anatomy.It is possible that the stent remains in the anatomy, as it could not be located within the devices.The procedure continued with implantation of a 3.5 x 16 mm non-abbott stent.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7741076
Report Number2024168-2018-05955
Device Sequence Number11025448
Product Code NIQ
UDI-Device Identifier08717648199523
UDI-Public08717648199523
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/18/2021
Device Catalogue Number1125350-15
Device Lot Number8013161
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2018
Initial Report FDA Received Date08/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age77 YR
-
-