PENUMBRA, INC. NEURON 6F 070 DELIVERY CATHETER; DQY Back to Search Results
Catalog Number PND6F070956M
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the neuron 070 was kinked approximately 8.0 cm from the hub.The neuron 070 was ovalized approximately 51.0, and 95.0 through 99.0 cm from the hub.Conclusions: evaluation of the returned neuron 070 revealed that the tip was ovalized.This type of damage typically occurs if the device is forcefully retracted out of its packaging tube at an extreme angle.Further evaluation of the neuron 070 revealed that the catheter shaft was kinked.This kink was likely incidental and may have occurred due to packaging of the device during the return to penumbra.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a medical procedure, the hospital staff found that the tip of the neuron 6f 070 delivery catheter (neuron 070) was ruined upon removal from the packaging.The damage to the neuron 070 was found prior to use.Therefore, the neuron 070 was not used in the procedure.The procedure was completed using a new neuron 070.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON 6F 070 DELIVERY CATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7741386
Report Number3005168196-2018-01500
Device Sequence Number1618524
Product Code DQY
UDI-Device Identifier00814548010076
UDI-Public00814548010076
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
K082290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 07/05/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/22/2021
Device Catalogue NumberPND6F070956M
Device Lot NumberF82054
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/05/2018
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date08/01/2018
Supplement Report FDA Received Date08/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1