BARD PERIPHERAL VASCULAR, INC. TRUE DILATATION CATHETER; BALLOON VALVULOPLASTY CATHETER Back to Search Results
Catalog Number 0264513
Medical Device Problem Codes Deflation Problem (1149); Retraction Problem (1536); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/05/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 04/28/2021), (b)(4).The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a valvuloplasty procedure, the ptv balloon allegedly ruptured and could not be deflated to fold into its original shape.It was further reported that there were alleged difficulties retracting the balloon from the patient, which led to detachment of the ptv balloon in the right common iliac artery.A snare was used to attempt to remove the detached balloon material, but was unsuccessful.The detached balloon material was left in the artery and a femoral-femoral bypass was done, as a result of the blocked artery.No further treatment was needed after the bypass procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE DILATATION CATHETER
Common Device Name
BALLOON VALVULOPLASTY CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7741390
Report Number2020394-2018-01339
Device Sequence Number16548985
Product Code OZT
UDI-Device Identifier00801741096631
UDI-Public(01)00801741096631
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K150667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0264513
Device Lot NumberGFCR1860
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2018
Initial Report FDA Received Date08/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
-
-