MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 9MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 9MM X 30CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Catalog Number M0036129300
Medical Device Problem Code Premature Separation (4045)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device is not available.
 
Event or Problem Description
It was reported that the physician attempted to retract the coil into the introducer sheath and the coil detached in the sheath.There was no clinical consequence to the patient as a result of this event.
 
Additional Manufacturer Narrative
The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.The subject device was not returned, therefore physical and functional tests could not be performed.Based on the information currently available an assignable cause for this event cannot be determined.
 
Event or Problem Description
It was reported that the physician attempted to retract the coil into the introducer sheath and the coil detached in the sheath.There was no clinical consequence to the patient as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 9MM X 30CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
MDR Report Key7741548
Report Number3008881809-2018-00330
Device Sequence Number13642433
Product Code HCG
UDI-Device Identifier04546540711618
UDI-Public04546540711618
Combination Product (Y/N)N
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Catalogue NumberM0036129300
Device Lot Number19847782
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/11/2018
Supplement Date Received by Manufacturer08/16/2018
Initial Report FDA Received Date08/01/2018
Supplement Report FDA Received Date08/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SL-10 MICROCATHETER (STRYKER)
-
-