• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD MYCARELINK; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD MYCARELINK; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number 24950G
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the remote monitor's reader was overheating.The remote monitor was power cycled.Troubleshooting steps were taken to no avail.The monitor is expected to be replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Product analysis: model: 24950g, serial/lot#: (b)(4): the remote monitor was returned and the analysis revealed that no complaint failure was found; found error code (b)(4) in failure analysis log.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The monitor was replaced and returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYCARELINK
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key7741718
Report Number3004593495-2018-00708
Device Sequence Number1589143
Product Code NVZ
UDI-Device Identifier00643169552104
UDI-Public00643169552104
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number24950G
Device Catalogue Number24950G
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Initial Date Received by Manufacturer 07/19/2018
Supplement Date Received by Manufacturer08/08/2018
02/21/2019
Initial Report FDA Received Date08/02/2018
Supplement Report FDA Received Date10/04/2018
03/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
-
-