| Model Number |
945ECLC |
| Medical Device Problem Codes |
Intermittent Continuity (1121); Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
A healthcare professional (hcp) reported that the controller was intermittently not working and shuts down in the middle of cases.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis found that the customer's complaint could not be duplicated.The power supply was loose due to incorrect screws.The screws were replaced and tightened.The device was cleaned, tested and passed according to manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|