• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC XOMED INC. NIM-ECLIPSE® CONTROLLER; STIMULATOR, ELECTRICAL, EVOKED RESPONSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC XOMED INC. NIM-ECLIPSE® CONTROLLER; STIMULATOR, ELECTRICAL, EVOKED RESPONSE Back to Search Results
Model Number 945ECLC
Medical Device Problem Codes Intermittent Continuity (1121); Use of Device Problem (1670)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare professional (hcp) reported that the controller was intermittently not working and shuts down in the middle of cases.There was no patient impact.
 
Additional Manufacturer Narrative
Analysis found that the customer's complaint could not be duplicated.The power supply was loose due to incorrect screws.The screws were replaced and tightened.The device was cleaned, tested and passed according to manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIM-ECLIPSE® CONTROLLER
Common Device Name
STIMULATOR, ELECTRICAL, EVOKED RESPONSE
Manufacturer (Section D)
MEDTRONIC XOMED INC.
6743 southpoint dr n
jacksonville FL 32216
MDR Report Key7741995
Report Number1045254-2018-00345
Device Sequence Number16663007
Product Code GWF
UDI-Device Identifier00643169071872
UDI-Public00643169071872
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K050798
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number945ECLC
Device Catalogue Number945ECLC
Device Lot Number212558426
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2018
Supplement Date Received by Manufacturer08/08/2018
09/19/2018
Initial Report FDA Received Date08/02/2018
Supplement Report FDA Received Date10/03/2018
10/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-