| Model Number |
X SERIES |
| Medical Device Problem Code |
Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device failed to discharge.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed during review of the device history logs.However, the reported problem could not be duplicated with the device.The defib receptacle was replaced as a precaution to reduce the probability of reoccurrences.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|