BECTON DICKINSON MEDICAL SYSTEMS BD LUER-LOK SYRINGE WITH PRECISIONGLIDE ¿ NEEDLE; SYRINGE WITH NEEDLE Back to Search Results
Catalog Number 309569
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd luer-lok syringe with precisionglide ¿ needle had twisted hubs causing the needle to bend when used, this could lead to possible leakage.There was no report of exposure, serious injury or medical intervention.
 
Event or Problem Description
It was reported that bd luer-lok syringe with precisionglide ¿ needle had twisted hubs causing the needle to bend when used, this could lead to possible leakage.There was no report of exposure, serious injury or medical intervention.
 
Additional Manufacturer Narrative
Investigation summary: one photo depicting a 3ml syringe with needle.The needle hub is twisted/damaged.The twisted hub/damage observed is rejectable per product specification.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.Investigation conclusion: potential root cause for the twisted hub defect is associated with the assembly process.No corrective actions are necessary based on the defective rate identified.Correction: due to an it issue beginning on 7/3/2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacture: no.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD LUER-LOK SYRINGE WITH PRECISIONGLIDE ¿ NEEDLE
Common Device Name
SYRINGE WITH NEEDLE
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
MDR Report Key7742185
Report Number1213809-2018-00481
Device Sequence Number16570465
Product Code FMF
UDI-Device Identifier30382903095699
UDI-Public30382903095699
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K980987
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2023
Device Catalogue Number309569
Device Lot Number8106769
Initial Date Received by Manufacturer 07/12/2018
Supplement Date Received by Manufacturer07/12/2018
Initial Report FDA Received Date08/02/2018
Supplement Report FDA Received Date08/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-