• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-522NAL
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that the insulin pump alarmed with a compromised force sensor system during the priming process.The patient's blood glucose at the time of incident is unknown.Troubleshooting was initiated for the rewind and prime issue.The drive support cap was flush.The customer was advised to discontinue use of the insulin pump and revert to a back-up plan per health care professional¿s instruction.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device did not have a battery installed when received.No compromised force sensor system alarm noted during testing.Device primed properly during the prime test, however device had a loose or flush drive support disk.Device passed the rewind, prime test, and excessive no delivery test.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and unexpected restart error test.Unable to perform the displacement test, basic occlusion test, occlusion test and delivery accuracy test due to broken reservoir tube lip.Device also had a minor scratched display window, cracked case at display window corner, cracked reservoir tube, cracked reservoir tube window and a missing end cap sticker.The test p-cap and reservoir will not lock in place in the reservoir compartment due to broken reservoir tube lip.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-522NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7743438
Report Number3004209178-2018-98914
Device Sequence Number1365466
Product Code OYC
UDI-Device Identifier00613994402110
UDI-Public(01)00613994402110
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-522NAL
Device Catalogue NumberMMT-522NAL
Device Lot NumberA5522NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2018
Initial Date Received by Manufacturer 07/05/2018
Supplement Date Received by Manufacturer02/04/2019
Initial Report FDA Received Date08/02/2018
Supplement Report FDA Received Date02/07/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age56 YR
Patient Weight127
-
-