| Model Number |
MMT-523NAH |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
05/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call the insulin pump had a keypad anomaly.The customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump down button was unresponsive.No significant events leading to keypad anomaly were observed.Keypad anomaly troubleshooting was performed and confirmed that time is advancing.Customer also reported that the insulin pump was squirting out during manual prime process.Customer had a high blood glucose of in the 400 mg/dl to 500 mg/dl and treated manual injection and insulin pump.Customer declined high blood glucose troubleshooting.The customer was advised to discontinue use of pump and revert to back-up plan.The insulin pump is being replaced and is expected to return for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Insulin pump received with intermittent button response due to corroded keypad traces.No button error alarm during testing.Insulin pump passed displacement test and self test.
|
| |
|