Model Number 46128
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/25/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported the system was placed on (b)(6) 2018.The tubing was found to be loose and disconnected near the injection site by itself.The system was replaced on (b)(6).There were no patient symptoms or complications associated with this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
The returned product was patent.The device did not meet the requirements for leak testing due to the patient line tubing-luer connection being disconnected at the ¿y¿ connector.Adhesive was present at the connection.It is unknown how or when the damage occurred.All edms devices are 100% leak tested at the time of manufacture.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BECKER EXTERNAL DRAINAGE AND MONITORING SYSTEM
Common Device Name
DEVICE, MONITORING, INTRACRANIAL PRESSURE
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key7744649
Report Number2021898-2018-00385
Device Sequence Number16563789
Product Code GWM
UDI-Device Identifier00673978228754
UDI-Public00673978228754
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K984053
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/11/2022
Device Model Number46128
Device Catalogue Number46128
Device Lot Number213999476
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2018
Supplement Date Received by Manufacturer08/08/2018
10/15/2018
Initial Report FDA Received Date08/02/2018
Supplement Report FDA Received Date10/02/2018
10/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age29 YR
-
-