| Model Number |
MZ9265 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Mz9271; non-bd picline and a 10ml bd syringe; td (b)(6) 2018.The product has been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
| |
|
Event or Problem Description
|
|
The customer reported that fluid leaked from above the luer lock on bi-fuse set.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report of leaking was confirmed.Visual inspection and functional testing showed no anomalies.Pressure testing showed a leak at the connection between the male luer of a concomitant extension set and the female port of the maxzero extension set.Closer inspection of the maxzero under magnification showed a microchannel that was allowing fluid to leak out.The root cause of the microchannel was an equipment error during the inspection and testing process.
|
| |
|
Event or Problem Description
|
|
The customer reported that fluid leaked from above the luer lock on bi-fuse set.There was no report of patient harm.
|
| |
|