Model Number ETBF2516C166E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii stent graft system was implanted in a patient for the endovascular treatment of a 63mm diameter abdominal aortic aneurysm.It was reported that during the index procedure a type ia endoleak was detected.The patient had a short 5mm neck with posterior bulge distal to 5mm neck.A total of 10 endoanchors were placed as part of the planned procedure due to the short proximal neck but the type ia endoleak persisted.The posterior bulge filled with contrast abruptly during angiogram but onlya small leak of contrast went into the aaa.Ct the following day showed a minimal type ia endoleak entering the sac and a larger type ii endoleak.Per the physician, the cause of the event is due to patients¿ anatomy with short neck with posterior bulge.No additional clinical sequelae were reported, and the patient will be monitored.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key7746058
Report Number9612164-2018-01944
Device Sequence Number1258536
Product Code MIH
UDI-Device Identifier00613994991270
UDI-Public00613994991270
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/22/2020
Device Model NumberETBF2516C166E
Device Catalogue NumberETBF2516C166E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/12/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date08/03/2018
Supplement Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age76 YR
-
-