| Model Number |
S53153 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Manufacturers initial investigation: device history record will be reviewed.The defect product was not returned to the manufacturer.The customer returned the remainder of the unopened needles.These will be evaluated.Justification for not providing below information and applicable sections: patient information - no patient injury reported, device malfunction occurred.Relevant tests / laboratory data - this section is not applicable as no patient injury occurred.Other relevant history, including preexisting medical conditions: this section is not applicable as no patient injury occurred.Serial # - this section is not applicable as the medical device does not have a serial number.Implantable.Reprocessor name and address - this section is not applicable as the medical device is not a single-use device that was reprocessed or reused on a patient.If remedial action initiated , check type - this section is not applicable as no remedial action was initiated.There was no action reported under 21usc 360i(f).
|
| |
|
Event or Problem Description
|
|
Hub and cannula separation from colour cover.A red emg needle came out of its protective cylinder while performing an emg, the needle came loose from the red housing on the s53153 teca concentric needle.
|
| |
|
Additional Manufacturer Narrative
|
|
Two subsequent requests were sent to the customer requesting product return and further information from the customer.No response has been received from the customer.The complaint log was reviewed over a period of 12 months.3 similar complaints were noted for this same issue.Capa (b)(4) has been opened to investigate this issue further.
|
| |
|
Additional Manufacturer Narrative
|
|
Work order (b)(4) was reviewed.No rework or sort performed on this lot.No excessive scrap quantities were noted.No ncr's noted.Accepted run at risk 179 - related to the refurbishment of the millipore system.Accepted run at risk 180 - related to retest of environmental monitoring of cleanroom.In process testing and inspections recorded on (b)(4) rev g were reviewed.All samples taken for tests gave acceptable readings.Hub to cover retention results were within specification minimum 1.2kgf.Lowest value recorded 1.840kgf, highest recorded value 6.190kgf.Hub to cannula tensile test spec minimum 6lbf.Lowest value recorded 14.28 lbf, highest recorded value 20.00 lbf.The following tests were carried out per (b)(4) rev n on a qty of 25 returned needles (unopened): hub to cover retention test- all needles passed with results within minimum spec of 1.2kgf.Lowest value recorded 1.805kgf, highest recorded value 4.385kgf.Hub to cannula tensile test - all needles passed with results within minimum spec of 6lbf.Lowest value recorded 16.12 lbf, highest recorded value 20.22 lbf.Risk management plan for teca concentric needles was reviewed.(b)(4) risk and risk management report - teca disposable needles and value disposable needles and value line dcn was reviewed.There are two potential hazards and hazardous situations associated with physical injury - needle stick, id 25 (needles are sharp, users could pierce themselves) and id 31 (cannula & hub assembly separates from outer color cover).The overall risk rating is 3a (severity of hazard - 3- moderate, probability of hazard - a - unlikely).
|
| |
|