Patient underwent a medtronic aicd placement in late (b)(6) 2018.The leads and generator were placed without any issues noted.Ventricular fibrillation was induced, but the patient could not be defibrillated with the device despite several different configurations.The ventricular lead was disconnected and repositioned along the interventricular septum and actively fixed in position.The new ventricular stimulation threshold measured 1 volt with a lead resistance of 471 ohms and an r-wave amplitude of 7.6 mv.The ventricular lead was secured to the pectoralis muscle with 2-0 silk suture and a 2-0 silk pursestring suture was then placed around the leads at the entry site beneath the clavicle.The lead was connected to the defibrillator which was returned to the pocket.Ventricular fibrillation was induced, but the patient could not be defibrillated with the device and required external defibrillation.A decision was made to leave the device in place at the current time and retest at a later time.The patient tolerated the procedure with no known harm.
|