| Model Number |
SEDR01 |
| Medical Device Problem Code |
Failure to Interrogate (1332)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/09/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Company rep.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable pulse generator (ipg) could not be fully interrogated, and it had reached elective replacement indicator (eri) status quite a while ago.In addition, lead integrity alerts had triggered, and the right atrial (ra) lead exhibited high impedance.Further follow-up investigation confirmed that the ra lead had fractured.The device was removed and replaced, and the ra lead was capped and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|