Catalog Number MPIS-501-NT-U-SST
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, during a device failure analysis of the micropuncture transitionless access set that a partial separation occurred.The original incident was noted as difficult advancement of the device in an unspecified procedure.After removing the first device, another device was used and no further issues were reported.However, after manufacturer evaluation, it was determined that the coils appeared "pushed away" from the solder connection between the mandrill and coil.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the dimensional verification, complaint history, device history record, documentation, drawing, instructions for use (ifu), manufacturing instructions, quality control, specifications, trends, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a partial separation, where the coils appeared pushed away from the solder connection between the mandril and coil.This separation was noted at length of approximately 33.4 cm from the proximal end of the device.At the partial separation, the needle was not able to pass over the wire guide.The length of the coil/mandril separation was approx.0.2 cm.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.Measures are being conducted to address this failure mode.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROPUNCTURE TRANSITIONLESS ACCESS SET
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7746805
Report Number1820334-2018-02335
Device Sequence Number1399589
Product Code DYB
UDI-Device Identifier00827002438712
UDI-Public(01)00827002438712(17)210313(10)8662836
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/13/2021
Device Catalogue NumberMPIS-501-NT-U-SST
Device Lot Number8662836
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2018
Initial Date Received by Manufacturer 07/30/2018
Supplement Date Received by Manufacturer08/03/2018
Initial Report FDA Received Date08/03/2018
Supplement Report FDA Received Date08/09/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-