Medical Device Problem Codes Device Displays Incorrect Message (2591); Fail-Safe Problem (2936)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
 
Event or Problem Description
On (b)(6) 2018, the reporter contacted animas, alleging that there was a call service 064 alarm.This complaint is being reported because the issue may result in a long-term cessation of insulin delivery if the user is unable to resolve the alarm.There was no allegation of an adverse event associated with this complaint.
 
Additional Manufacturer Narrative
Device evaluation: the device has been returned and evaluated by product analysis on 08-oct-2018 with the following findings: a review of the pump¿s black box and alarm history showed a cs 064 call service.During testing, the pump was powered on to the verify screen with no alarms.The pump rewind, load and prime steps were performed successfully and the pump was exercised for 12 hours with no call service alarms occurring.The complaint of a cs 064 call service alarm issue was not duplicated during investigation.Unrelated to the complaint, the battery compartment was observed to be cracked below the bumper pad.The pump was opened and there was evidence of moisture corrosion on ir and on transceiver boards, and near the motor flex connector.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANIMAS VIBE
Common Device Name
INSULIN INFUSION PUMP
Manufacturer (Section D)
ANIMAS LLC
200 lawrence dr
west chester PA 19380 3428
MDR Report Key7746872
Report Number2531779-2018-14132
Device Sequence Number11201771
Product Code OYC
UDI-Device Identifier10840406100549
UDI-Public10840406100549
Combination Product (Y/N)Y
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P130007
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/03/2018
Was the Report Sent to FDA? Yes
Device Age9 MO
Initial Date Received by Manufacturer 07/12/2018
Supplement Date Received by Manufacturer07/12/2018
Initial Report FDA Received Date08/03/2018
Supplement Report FDA Received Date10/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age8 YR
-
-