Model Number FGS-0313
Device Problems Loss of or Failure to Bond (1068); Failure to Fire (2610)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 07/02/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, they had a capsule which failed to attach, the capsule did not deploy.There was no harm to the patient, no intervention was required, and they tried to repeat the procedure, but it also failed.There was nothing unusual about the patient or the procedure and an endoscopy had been performed prior to the procedure and showed the esophagus to be normal.No lubrication was used to facilitate placement of the capsule, the capsule has been cleaned and decontaminated, and the delivery system and capsule will be returned for investigation.
 
Manufacturer Narrative
Corrected information: if information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional: evaluation summary: this report is based on information provided by medtronic investigation personnel and the sample that arrived.One bravo delivery device were received for evaluation.A review of the product expiration date discovered this product was used before the expiration date.The returned sample met specification as received by medtronic.The visual inspection found no notable conditions.The customer reported bravo failed to attach to patient's esophagus.The investigation of the returned equipment did not identify anything that would have caused or contributed to the reported event.The investigation found the device to function normally and within specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer Contact
amy beeman
5920 longbow drive
boulder, CO 80301
7632104064
MDR Report Key7747065
MDR Text Key116237946
Report Number9710107-2018-00882
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101361688
UDI-Public7290101361688
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-0313
Device Catalogue NumberFGS-0313
Device Lot Number38210Q
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/17/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-