• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS NAVARRE PERCUTANEOUS BILIARY DRAINAGE CATHETER WITH NITINOL, 12F LOCKING PIGTAIL; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS NAVARRE PERCUTANEOUS BILIARY DRAINAGE CATHETER WITH NITINOL, 12F LOCKING PIGTAIL; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number NBD12
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that as the nurse was inserting the stylet (before inserting the catheter over the guidewire), it was observed that there was a tear at the tip of the catheter.The nurse had been prepping the patient for a biliary drain procedure.A new tube was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample returned was one 12 f locking pigtail navarre catheter with flexible cannula installed and a photograph of the tip of a locking pigtail navarre catheter which is consistent with the returned physical sample.The metal cannula was also returned.Detailed examination of the failure and verification of the component lengths was performed.The investigation is confirmed for (b(4) - subq leak/break/split w/o embo as the tip was found to be fractured.The definitive root cause could not be determined based upon available information.It is unknown if procedural issues contributed to the reported event.Labeling review: a review of the product instructions for use (ifu) procedural instructions, indications, warnings, precautions, cautions, possible complications and contraindications showed that the product labeling is adequate.
 
Event or Problem Description
It was reported that as the nurse was inserting the stylet (before inserting the catheter over the guidewire), it was observed that there was a tear at the tip of the catheter.The nurse had been prepping the patient for a biliary drain procedure.A new tube was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVARRE PERCUTANEOUS BILIARY DRAINAGE CATHETER WITH NITINOL, 12F LOCKING PIGTAIL
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7747139
Report Number3006260740-2018-01942
Device Sequence Number13843228
Product Code FGE
UDI-Device Identifier00801741036859
UDI-Public(01)00801741036859
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K955110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNBD12
Device Catalogue NumberNBD12
Device Lot NumberGFBX3913
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/03/2018
Initial Date Received by Manufacturer 07/10/2018
Supplement Date Received by Manufacturer09/17/2018
Initial Report FDA Received Date08/03/2018
Supplement Report FDA Received Date10/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-