| Model Number |
E14554NSB |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Component Missing (2306); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
07/11/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the device was felt the insulation was loose.The insulation broke in pieces while using, and some of the pieces came off.One of the fallen pieces was missing and could not be found as it fell into patient's cavity and could not be retrieved.Also the physician tore a part of the insulation.There was no patient harm.A photo was submitted with markings as encircled 1 and 2 indicate the pieces came off, and encircled 1 went missing, encircled 3 indicates the piece was torn by the physician.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: the product was not returned; however, a picture was provided by the customer for analysis.The product in the picture did not meet specification.The picture showed that the blue heat shrink to have been scratched and damaged allowing the polytetrafluoroethylene (ptfe) insulator to get torn.The reported condition of insulation was confirmed, whereas the other three conditions could not be confirmed.The received picture shows that the blue insulation and the grayish ptfe insulator had damaged and torn.The blue heat shrink insulation holds the ptfe in place and damage to the insulation can cause the clear insulation to be torn.The damage to the insulation indicates the user mishandled the electrodes.One of the possible reasons is cleaning with an abrasive material.This cleaning method would also contribute to the insulator disengaging.The investigation identified the root cause of the reported event to be attributed to user handling damage.The instructions for use (ifu) states: the electrode has a coating to reduce sticking of eschar.Cleaning the electrode with a scratch pad or other abrasive object, scraping with a sharp object or bending beyond 90 degrees may damage the electrode.If the electrode is damaged, discard it.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: two devices were received for evaluation.Device 1 did not meet the specification, whereas device 2 (which was unused) met specification as received.Visual inspection of device #1 (incident sample) showed that the blue heat shrink to have been scratched and damaged allowing the ptfe insulator to get torn.Device #2 was unused and did not show any defects.The customer reported that the device component disengaged into cavity (and not retrieved), the component was missing, the device fragments were in patient and the insulation was damaged.The reported condition of insulation was confirmed, whereas the other three conditions could not be confirmed.The investigation showed that the blue insulation and the grayish ptfe insulator had damaged and torn.The blue heat shrink insulation holds the ptfe in place and damage to the insulation can cause the clear insulation to be torn.The loose ptfe insulator was returned.The damage to the insulation indicates the user mishandled the electrodes.One of the possible reasons is cleaning with an abrasive material.This cleaning method would also contribute to the insulator disengaging.The investigation identified the root cause of the reported event to be attributed to user handling damage.The instructions for use (ifu) states: the electrode has a coating to reduce sticking of eschar.Cleaning the electrode with a scratch pad or other abrasive object, scraping with a sharp object or bending beyond 90 degrees may damage the electrode.If the electrode is damaged, discard it.If information is provided in the future, a supplemental report will be issued.
|
| |
|