DAVOL INC., SUB. C.R. BARD, INC. PHASIX ST MESH; SURGICAL MESH Back to Search Results
Catalog Number 1200015
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abdominal Pain (1685); Pain (1994); Hernia (2240)
Date of Event 07/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Based on the information provided we are unable to determine to what extent, if any the phasix st mesh may have caused or contributed to the patients hernia recurrence.Based on the available information, no conclusion can be made.Recurrence is a known inherent risk of hernia repair surgery.Regarding recurrence the warning section of the instructions-for-use state, "to prevent recurrences when repairing hernias, phasix st mesh must be large enough to provide sufficient overlap beyond the margins of the repair/primary closure.Careful attention to mesh fixation placement and spacing will help prevent excessive tension or gap formation between the mesh and fascial tissue.A review of the manufacturing records was performed and found that the lot was manufactured to specification.If additional information is provided, a supplemental emdr will be submitted.Note: the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.Remains implanted.
 
Event or Problem Description
It was reported to davol that a patient who is part of a clinical study experienced a hernia recurrence.On (b)(6) 2017 - the patient underwent a laparoscopic, intra-abdominal hernia repair and was implanted with a phasix st mesh.The phasix st is noted to have been positioned so that its edges extend beyond the margins of the defect by at least 5cm.On (b)(6) 2018 - the patient was reported to have had low quadrant abdominal pain.On (b)(6) 2018 - the patient presented for his 18 month follow up visit.A hernia recurrence was confirmed upon physical exam.This adverse event was assessed per the clinician as moderate in severity, possibly related to the study device and possibly related to the procedure.The adverse event is identified as not recovered / not resolved.No intervention has been taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PHASIX ST MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC.
100 crossings blvd.
warwick RI 02886
MDR Report Key7747252
Report Number1213643-2018-02597
Device Sequence Number1084393
Product Code OWT
UDI-Device Identifier00801741095290
UDI-Public(01)00801741095290
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K143380
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source health professional,other,stu
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/28/2018
Device Catalogue Number1200015
Device Lot NumberHUAU0482
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2018
Initial Report FDA Received Date08/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-