| Model Number |
60728818 |
| Medical Device Problem Codes |
Device Contamination with Chemical or Other Material (2944); Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/03/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Bayer product analysis received and examined one returned low pressure connector tubing (lpct).Visual examination confirmed the presence of a particle within the tube set.The particle was identified as degraded resin from the tubing extrusion process and was partially connected to the inside wall of the tubing.As the particle was partially embedded into the tubing wall, there is a potential that it could fragment while under pressures typically generated during an injection.Our investigation determined that the particle noted in the lpct was introduced during the manufacturing process and was not detected upon receipt of the product from the supplier.A 12 month review of complaint history determined the complaint rate to be (b)(4) complaints per million (cpm).
|
| |
|
Event or Problem Description
|
|
The customer reported the following: after opening the stellent ct disposable kit and upon inspection, they saw a foreign body within the low pressure connector tubing (lpct).The customer elected not to use the tubing.No injury or adverse event was reported.
|
| |
|
Search Alerts/Recalls
|