Catalog Number PNML6F088804M
Medical Device Problem Codes Break (1069); Device Damaged Prior to Use (2284)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/05/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the neuron max was partially fractured approximately 82.5 cm from the hub.The neuron max was damaged and, therefore, could not be functionally tested.Conclusions: evaluation of the returned neuron max revealed that the distal tip was partially fractured.If the packaging mandrel becomes unsecured from the packaging card it may pin the neuron max distal tip between the mandrel and packaging tube.If the neuron max is subsequently retracted while pinned, the distal tip will likely fracture.There was no visible damage to the dilator.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a medical procedure, the hospital staff found that the tip of the neuron max 6f 088 long sheath (neuron max) had a burr on it and the inner material was visible upon unpacking and flushing the neuron max.The damage to the neuron max was found prior to use.Therefore, the neuron max was not used in the procedure.The procedure was completed using a new neuron max.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX 6F 088 LONG SHEATH
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7747705
Report Number3005168196-2018-01483
Device Sequence Number1280180
Product Code DQY
UDI-Device Identifier00814548012209
UDI-Public00814548012209
Combination Product (Y/N)Y
Initial Reporter CountryNL
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/05/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/29/2021
Device Catalogue NumberPNML6F088804M
Device Lot NumberF81672
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/25/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2018
Initial Report FDA Received Date08/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-